# Mobility by impact excitation (ISO 7626-5)

An instrumented hammer is the quickest way to measure a mobility: no exciter
to attach, no drive rod to align, and the excitation point moves from one
location to the next in seconds. **ISO 7626-5:2019** (*Mechanical vibration and
shock — Experimental determination of mechanical mobility — Part 5:
Measurements using impact excitation with an exciter which is not attached to
the structure*) is the part of the ISO 7626 series that makes that measurement
accurate. It measures the same
frequency-response functions as
[ISO 7626-2](https://docs.jmrp.io/phonometry/vibration/structural/mechanical-mobility/), but the
record it works on has a peculiar shape: a force pulse a few milliseconds long,
followed by the free decay it leaves behind. Almost every clause of the part is
about a consequence of that shape. The pulse fills a tiny fraction of the
record, so the noise around it counts (8.5.1). The decay may not be over when
the record ends, so the spectrum leaks (8.3, 8.5.2). A second impact inside
the record cuts notches in the force spectrum (6.4). And the average runs over
impacts, not over segments of one long signal (8.6).

The part prints no worked example. The examples below strike a resonator whose
answer is known, 2 kg on a spring tuned to 50 Hz with 0.5 % of critical
damping, whose driving-point mobility peaks at $1/c = 0.159$ m/(N·s), and every
number they print can be checked against that.

## How the measurement goes

The hammer carries a force transducer between its mass and an interchangeable
tip (6.1). The structure hangs on a soft suspension or sits on its intended
support (clause 5), an accelerometer picks up the response, and a two-channel
Fourier analyser captures each impact with a little pre-trigger data, so the
leading edge of the pulse is not lost (8.2).

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/diagram_mobility_rig_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/diagram_mobility_rig.svg" alt="ISO 7626 mobility measurement: a free-free beam on soft suspension driven by an exciter through a drive rod and an impedance head at the driving point, an accelerometer at a transfer point, and an impact hammer as the alternative excitation. Below, a three-panel inset shows the accelerometer routed through the drive rod marked invalid, the force transducer at the structure marked valid, and the force transducer at the exciter end marked with caution, with the ten-times mobility criteria for the exciter attachment and the suspension stated underneath." width="92%"></picture>

The order of work the part implies, and the function that judges each step:

1. **Calibrate on a rigid block** of known mass, freely suspended, at the start
   and end of each series and whenever the tip or the mass changes (7.2). Its
   accelerance should read $1/m$ within ±5 %: `rigid_mass_calibration_check`,
   the ISO 7626-2 check that 7.2 calls essentially the same procedure.
2. **Check the channels against each other** by feeding one broad-band signal
   to both: unity within ±5 % and zero phase within ±5° (8.1),
   `verify_channel_match`.
3. **Set the frequency range and the record.** The sample rate is 2.56 to 4
   times the filter cut-off (8.3), and the record should be long enough for
   the response to decay to about 1 % of its start.
4. **Strike once at the highest force the test will use** and check the
   channels for saturation (8.4, `check_overload`) and the force spectrum
   across the range of interest (6.3, `check_force_spectrum`). Change the tip
   or the mass if the pulse is too short or too long.
5. **Strike at least five times** at the same point, letting the structure
   come to rest in between (8.6), and discard any record with a double hit
   (6.4, `check_double_hit`). Three to five impacts are usually enough to
   verify data quality in a low-noise environment (8.6), and five is the
   fewest records a high coherence is trusted from (9.1), which is what the
   coherence check of step 7 asks for.
6. **Judge the decay** (8.3, 8.5.2, `check_response_decay`, given the force
   so the impact itself is left out), and if it is too slow, add an
   exponential window and plan to correct for it (Annex A).
7. **Average** (8.6, `impact_mobility`) and **judge the coherence** (9.1,
   `check_coherence`).

## 1. What an impact record holds

The analyser samples $N$ points at an interval $\Delta t$, a record of length
$T = N\,\Delta t$ whose discrete Fourier transform has $N/2 + 1$ lines $1/T$
apart (8.3). When the transient lies wholly inside the record, the transform
scaled by $\Delta t$ samples the continuous Fourier transform
$F(f)$, and the part describes the pulse by its **energy spectral density**
(3.3, 3.4), the power spectral density $2\lvert F\rvert^2/T$ multiplied by the
record length:

$$
G_{FF}(f) = 2\,\lvert F(f)\rvert^2 \qquad [\mathrm{N^2\,s/Hz}].
$$

A pulse has a main lobe at low frequency and side lobes that fall quickly, and
its usable bandwidth is inversely proportional to its duration (6.2): a harder
tip or a lighter hammer widens it, a softer tip or a heavier hammer narrows it
(6.3).

The force typically occupies less than 1 % of the record (8.5.1), and
whatever noise the force channel carries runs through all of it. Because the noise adds to the
force auto-spectrum, averaging does not remove it. The **force window** does:
unity over the part of the record that holds the pulse and its filter
response, exactly zero everywhere else. The noise that reaches the spectrum is
then only the noise inside the window.

```python
import numpy as np
from scipy import signal
from phonometry import vibration

fs, n = 4096.0, 4096                      # a 1 s record: 1 Hz between lines
t = np.arange(n) / fs

# A 2 kg resonator tuned to 50 Hz with 0.5 % damping, struck by a 0.5 ms pulse.
m, fn, zeta = 2.0, 50.0, 0.005
wn = 2.0 * np.pi * fn
accelerance = signal.lti([1.0 / m, 0.0, 0.0], [1.0, 2.0 * zeta * wn, wn**2])
force = 100.0 * np.exp(-((t - 0.005) ** 2) / (2.0 * 0.0005**2))
_, accel, _ = signal.lsim(accelerance, force, t)

# The same pulse with broad-band noise on the force channel, and a force
# window that keeps the first 15 ms of the record (8.5.1).
rng = np.random.default_rng(7626)
noisy = force + 0.3 * rng.standard_normal(n)
window = vibration.force_window(n, fs, width_s=0.015)
f, whole = vibration.energy_spectral_density(noisy, fs)
_, kept = vibration.energy_spectral_density(noisy * window, fs)
above = f > 1500.0                        # where the pulse itself has gone
print(round(10 * np.log10(whole[above].mean() / kept[above].mean()), 1))  # 18.4 dB
print(round(10 * np.log10(1.0 / 0.015), 1))    # 18.2 dB, 10 lg(T / width)
```

The floor drops by the ratio of the record to the window, 18 dB here, and the
pulse is untouched because the window never attenuates it. Two warnings come
with the window. A rectangular one spreads periodic noise and DC offset over a
wide band (8.5.1, Figure 8), so remove those before windowing; `taper_s` adds
half-cosine ramps outside the unity part if a smooth edge is wanted. And it
must **never** cut a second impact out of the record (6.4): the response still
carries that impact. `impact_mobility` warns with `ImpactExcitationWarning`
when its force window zeroes an impact the record holds.

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/impact_force_window_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/impact_force_window.svg" alt="Two panels. Left, the first 40 milliseconds of a force record: a Gaussian pulse of about 100 newtons at 5 milliseconds on a low noise floor, and a dashed force window at unity from 0 to 15 milliseconds and zero after it. Right, the force energy spectral density on a logarithmic scale up to 2 kilohertz: the whole record flattens into a noise floor near ten to the minus four from about 800 hertz upwards, the windowed record follows the pulse alone down past ten to the minus six before its own floor, and the pulse alone falls smoothly." width="92%"></picture>

*The force window at work. The pulse fills a few samples of the one-second
record; everything after 15 ms is noise, and zeroing it lowers the floor of the
force spectrum by about 18 dB without changing the pulse's own spectrum, which
is what 8.5.1 means by a window that "introduces no distortion".*

<details>
<summary>Show the code for this figure</summary>

The force spectrum of the windowed record, read against a frequency range of
interest, is what `ForceSpectrumCheck.plot()` draws:

```python
vibration.check_force_spectrum(noisy * window, fs, frequency_range_hz=(0.0, 800.0),
                               max_drop_db=20.0).plot()
```

The figure itself, by hand:

```python
import matplotlib.pyplot as plt

fig, (left, right) = plt.subplots(1, 2, figsize=(12.4, 5.2))
shown = t <= 0.04
left.plot(1e3 * t[shown], noisy[shown], label="force record with noise")
left.plot(1e3 * t[shown], 110.0 * window[shown], "--",
          label="force window (unity for 15 ms)")
left.set_xlabel("Time [ms]"); left.set_ylabel("Force [N]"); left.legend()

_, clean = vibration.energy_spectral_density(force, fs)
band = (f > 0.0) & (f <= 2000.0)
right.semilogy(f[band], whole[band], color="0.6", label="whole record")
right.semilogy(f[band], kept[band], label="force window")
right.semilogy(f[band], clean[band], "--", label="the pulse alone")
right.set_ylim(1e-7, 1.0)
right.set_xlabel("Frequency [Hz]"); right.set_ylabel("$G_{FF}$ [N²·s/Hz]")
right.legend(); plt.show()
```

</details>

## 2. The averaged estimate and its coherence (8.6, 9.1)

Averaging over several impacts at one point improves the estimate, and the
part prescribes the form: the **averaged cross-spectrum** of response and force
divided by the **averaged auto-spectrum** of the force (8.6), the H1
estimator, with the ordinary coherence beside it (9.1):

$$
H = \frac{\sum_i X_i F_i^{*}}{\sum_i \lvert F_i\rvert^2}, \qquad
\gamma^2 = \frac{\left\lvert \sum_i X_i F_i^{*} \right\rvert^2}
               {\sum_i \lvert F_i\rvert^2 \,\sum_i \lvert X_i\rvert^2}.
$$

In a low-noise environment three to five impacts are usually enough to verify
the data (8.6), and the response must have decayed below the limit of detection
before the next impact, or its ringing corrupts the next record.

```python
# Five impacts of different strength, a little noise on the accelerometer.
gains = np.array([0.8, 1.0, 1.1, 0.9, 1.2])[:, np.newaxis]
forces = force * gains
accels = accel * gains + 0.02 * rng.standard_normal((5, n))
estimate = vibration.impact_mobility(forces, accels, fs, force_window_s=0.015,
                                     frequency_range_hz=(5.0, 800.0))
print(estimate.impacts, estimate.measured_frf)     # 5 accelerance

coh = vibration.check_coherence(estimate, frequency_range_hz=(10.0, 400.0))
print(coh.passes, round(float(coh.coherence.min()), 3))   # True 0.992
print(round(float(np.nanmax(coh.random_error_percent)), 2))  # 2.83 %
```

Each row of `forces` and `accels` is one impact. The response channel measured
acceleration, so the estimate is formed as an accelerance and converted to
mobility (`estimate.to(...)` gives any other kind). `frequency_range_hz` keeps
it to the frequency range of interest of 3.2: above the usable bandwidth of the
pulse the spectrum is noise over almost no force, and nothing built on the
result, its plot, its fiche or a fit, should see that part.

Records read from a measurement file arrive as a
[`Signal`](https://docs.jmrp.io/phonometry/io/audio-files/): pass it in place of the array
and leave out `fs`, which it brings with it (one impact per channel). A force
and an acceleration are not pressures, so a calibration factor the `Signal`
carries is not applied to them. The checks of section 5 that read a record
take a `Signal` the same way.

The coherence check passes when $\gamma^2$ exceeds **0.9** at every frequency it
judges and at least **five** impacts were averaged. Both numbers are 9.1's: a
coherence above 0.9 is what it calls high, and five to ten records are what a
high coherence needs for confidence in its own estimate. Five is also the top
of the three to five that 8.6 finds usually enough, so averaging five satisfies
both; `minimum_records` lowers the count when fewer were judged enough. A
single impact has a coherence of 1 by construction, which is why it fails the
second condition.
Low coherence at an antiresonance is not generally a concern (9.1 NOTE 1), and
`exclude_hz=[(f_low, f_high), ...]` leaves such bands out of the verdict. The
coherence does not see everything: a repeatable impact hides leakage, structural
nonlinearity and clipping (9.1 NOTE 2), which is what the checks of section 5
are for.

## 3. Leakage, and the exponential window (8.3, 8.5.2)

The estimate above has a flaw the coherence cannot see:

```python
at_50 = estimate.frequencies == 50.0
print(round(float(estimate.magnitude[at_50][0]), 4))   # 0.1257 m/(N·s)
print(round(1.0 / (2.0 * zeta * m * wn), 4))           # 0.1592, the true peak
```

The peak is 21 % low. The resonator is so lightly damped that it is still
ringing when the one-second record ends, and the transform of a truncated
decay leaks. The part's rule is that the response should decay to about **1 %**
of its initial magnitude by the end of the record, a figure 8.3 calls a
compromise: a longer record improves the frequency resolution but can lower
the signal-to-noise ratio. It offers a convenient check too: the peak at the
midpoint of the record about **10 %** of the highest (8.3, 8.5.2).

```python
# Give the check the force of the same impact: a driving-point acceleration
# follows the force during the impact (the mass line), so the decay is judged
# from the end of the pulse on.
decay = vibration.check_response_decay(accel, fs, force=force)
print(decay.passes, round(100 * decay.end_ratio, 1))    # False 21.1 (%)
print(round(100 * decay.midpoint_ratio, 1))             # 46.3 (%), not 10 %
```

The level at the end is read at the last sample. The largest value over the
last segment of the record sits at the segment's start, above the end of a
decay still under way, so the check carries it to the last sample at the rate
the peaks fall from the midpoint to the last segment, the steady decay the
midpoint check itself assumes. The verdict compares that level with the
printed figure to a whole per cent, so "about 1 %" passes any level below
1.5 %, which rounds to 1 %. Without the force, the highest peak of this record
is the impact, two and a half times the free decay, and the same record would
read 8.5 %.

A longer record would need a block size or a zoom band the analyser may not
have (8.3). The alternative the part recommends is an **exponential window**,
$w(t) = \mathrm{e}^{-at}$, which starts at unity and adds a known decay to the data so
the record ends where it should. As a general guideline, the natural decay
should still reach **25 %** or less within the record, or the correction of
Annex A becomes too sensitive to errors in the damping estimate (8.5.2), and
the check applies that limit when told a window will be used:

```python
windowed_decay = vibration.check_response_decay(accel, fs, force=force,
                                                exponential_window=True)
print(windowed_decay.passes, windowed_decay.limit_ratio)   # True 0.25

rate = vibration.exponential_decay_rate(n, fs, final_value=0.01)
print(round(rate, 3))                                      # 4.606 1/s
windowed = vibration.impact_mobility(forces, accels, fs, force_window_s=0.015,
                                     exponential_decay_rate_per_s=rate,
                                     frequency_range_hz=(5.0, 800.0))
print(round(float(windowed.magnitude[at_50][0]), 4))       # 0.0404 m/(N·s)
```

The window traded one error for another. The leakage is gone, but the peak is
now a quarter of the true one: the window multiplies the whole response by a
decay, and a mode that decays faster looks more damped. That is not a mistake
to be avoided but a known quantity to be taken out, which is what Annex A does.

## 4. Taking the window out again (Annex A)

Multiply both sides of the convolution $x(t) = \int_0^t h(\tau) f(t - \tau)\,
\mathrm{d}\tau$ (Formula (A.1)) by $\mathrm{e}^{-at}$, and because
$\mathrm{e}^{-at} = \mathrm{e}^{-a\tau} \mathrm{e}^{-a(t-\tau)}$ the windowed response is the windowed
force convolved with a windowed impulse response (Formula (A.2)). Every mode
$A_r \mathrm{e}^{s_r t}$ of the impulse response becomes $A_r \mathrm{e}^{(s_r - a)t}$: the
window moves each pole back by $a$ and leaves its residue alone. A damping
ratio $\hat\zeta_r$ estimated from windowed data therefore carries the
window's share, and Formula (A.3) removes it:

$$
\zeta_r = \hat{\zeta}_r - \frac{a}{\omega_r},
$$

with $\omega_r$ the damped natural frequency in rad/s. For a lightly damped,
well separated mode the measured peak is multiplied by
$\hat\zeta_r/\zeta_r$ to give the true one.

The annex leaves the damping estimator open ("a variety of curve-fitting
methods"), and `single_mode_fit` is the one this library offers: a
rational-fraction fit of one mode over a band, written as a pole, its residue
and a constant for the modes outside the band. `ImpactMobilityResult.fit_mode`
runs it on the kind the response channel measured, the record the window
actually multiplied, and remembers the window's rate:

```python
fit = windowed.fit_mode((40.0, 60.0))
print(round(fit.damped_natural_frequency_hz, 2))             # 50.0 Hz
print(round(100 * fit.apparent_damping_ratio, 3))            # 1.966 %

correction = fit.correction
print(round(100 * float(correction.window_damping_ratio[0]), 3))  # 1.466 %, a / omega_r
print(round(100 * float(correction.damping_ratio[0]), 3))         # 0.499 %, Formula (A.3)
print(round(100 * float(correction.exact_damping_ratio[0]), 3))   # 0.5 %, the pole moved back
print(round(float(correction.peak_correction_factor[0]), 2))      # 3.94

f_d = fn * np.sqrt(1.0 - zeta**2)
print(round(abs(complex(fit.corrected_mobility(f_d))), 4))   # 0.1588 m/(N·s)
```

The damping comes back as 0.499 % against a true 0.5 %, and the synthesized
mobility, the fitted mode with its pole moved back by $a$, peaks at 0.1588
against the true 0.1592. What is left, about 0.2 %, is truncation again: the
windowed record still ends at about 0.2 % of its start, 21 % of natural decay
times 1 % of window.

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/impact_exponential_window_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/impact_exponential_window.svg" alt="Two panels. Left, the acceleration of the resonator over one second: the recorded response still rings at about a quarter of its initial amplitude at the end, the windowed response decays to almost nothing, and a dashed exponential window falls from the initial amplitude to one per cent of it. Right, the mobility magnitude from 40 to 60 hertz on a logarithmic scale: the resonator peaks at about 0.16, the unwindowed estimate falls short of it at the peak and runs low on both flanks, the exponentially windowed estimate is smooth but peaks at only 0.04, and the mode with the window taken out by Annex A lies on the resonator's curve." width="92%"></picture>

*What the window takes and what Annex A gives back. Unwindowed, the estimate
leaks: it is low at the peak and low on the flanks by an amount no correction
can predict (8.5.2). Windowed, it is smooth but too damped, by an amount that
is known, and moving the fitted pole back by $a$ lands on the resonator.*

<details>
<summary>Show the code for this figure</summary>

`SingleModeFitResult.plot()` draws the data fitted, the fitted mode and the
mode with the window taken out:

```python
fit.plot()
```

The figure itself, by hand:

```python
import matplotlib.pyplot as plt

fig, (left, right) = plt.subplots(1, 2, figsize=(12.4, 5.4))
w = vibration.exponential_window(n, fs, decay_rate_per_s=rate)
free = t >= 0.02
left.plot(t[free], accel[free], color="0.6", lw=0.6, label="response as recorded")
left.plot(t[free], (accel * w)[free], lw=0.6, label="times the exponential window")
left.plot(t, float(np.max(np.abs(accel[free]))) * w, "--", label="window, 1 % at the end")
left.set_xlabel("Time [s]"); left.set_ylabel("Acceleration [m/s²]"); left.legend()

single = vibration.impact_mobility(force, accel, fs, frequency_range_hz=(5.0, 800.0))
single_windowed = vibration.impact_mobility(force, accel, fs,
                                            exponential_decay_rate_per_s=rate,
                                            frequency_range_hz=(5.0, 800.0))
dense = np.linspace(40.0, 60.0, 2001)
band = (single.frequencies >= 40.0) & (single.frequencies <= 60.0)
exact = vibration.sdof_mobility(dense, m, m * wn**2, 2.0 * zeta * m * wn)
right.semilogy(dense, np.abs(exact), "k:", label="the resonator")
right.semilogy(single.frequencies[band], single.magnitude[band], "o-", color="0.6",
               label="no window: truncated, leaking")
right.semilogy(single_windowed.frequencies[band], single_windowed.magnitude[band],
               "o-", label="exponential window: smooth, too damped")
single_fit = single_windowed.fit_mode((40.0, 60.0))
right.semilogy(dense, np.abs(single_fit.corrected_mobility(dense)), "--",
               label="window taken out by Annex A")
right.set_xlabel("Frequency [Hz]"); right.set_ylabel("Mobility |Y| [m/(N·s)]")
right.legend(); plt.show()
```

</details>

### Formula (A.3) is the first order of the pole shift

The annex states the effect exactly, the pole replaced by $s_r - a$, and then
gives (A.3) for the damping. The two are not the same thing. The windowed
pole has the decay rate $\hat\sigma_r = \sigma_r + a$ at the same damped
frequency $\omega_r$, and a damping ratio is a decay rate over the *undamped*
frequency, $\zeta = \sigma/\sqrt{\sigma^2 + \omega_r^2}$, which moves with the
decay rate. Solving that exactly and expanding gives (A.3) plus a remainder of
about $(\hat\zeta_r^3 - \zeta_r^3)/2$, always on the low side. The result
carries both: `damping_ratio` is (A.3) as printed and `exact_damping_ratio` is
the damping of the pole moved back, which is also what `corrected_frf` and
`corrected_mobility` synthesize.

The difference is invisible with the window of the example, which adds 1.5 %
of damping, and grows quickly with a heavier one:

```python
heavy = vibration.impact_mobility(forces, accels, fs, force_window_s=0.015,
                                  exponential_decay_rate_per_s=20.0,
                                  frequency_range_hz=(5.0, 800.0))
heavy_correction = heavy.fit_mode((40.0, 60.0)).correction
print(round(100 * float(heavy_correction.apparent_damping_ratio[0]), 3))  # 6.851 %
print(round(100 * float(heavy_correction.damping_ratio[0]), 3))           # 0.484 %
print(round(100 * float(heavy_correction.exact_damping_ratio[0]), 3))     # 0.5 %
```

This is not a misprint. The annex writes (A.3) for a "damping factor" it does
not define, and the part takes its terms from ISO 2041 (clause 3), whose
damping ratio is the damping coefficient over the critical one, $c/c_c$
(ISO 2041:2018, 3.2.96). For that ratio (A.3) is the first order of the pole
shift; for a decay rate over the damped frequency, $\sigma_r/\omega_r$, it
would be exact. It is a reason to keep the window light, the same advice 8.5.2
gives for another reason, and to prefer the exact form when the window adds
several per cent.

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/impact_a3_first_order_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/impact_a3_first_order.svg" alt="The error of the corrected damping, in per cent of the true damping, against the damping the exponential window adds from 1 to 10 per cent. Four measured curves, one per true damping of 0.2, 0.5, 1 and 2 per cent, stay near zero while the window adds less than about 2 per cent and then fall ever more steeply, the lightest damping fastest, to minus 26 per cent at a window of 10 per cent; the curves for the exact pole shift lie along zero at every window. A vertical dashed line at 1.5 per cent marks the window of this guide." width="82%"></picture>

*Formula (A.3) against the exact pole shift, each point a full measurement:
the resonator struck, windowed, fitted and corrected. The lighter the
structure, the sooner (A.3) falls behind, because the window's damping then
dominates the apparent damping the formula starts from.*

<details>
<summary>Show the code for this figure</summary>

`ExponentialWindowCorrection.plot()` splits each mode's apparent damping into
the true part and the window's:

```python
fit.correction.plot()
```

The figure itself, by hand, one measurement per point:

```python
import matplotlib.pyplot as plt

ratios = np.linspace(0.01, 0.1, 19)
omega_d = 2.0 * np.pi * fn
for true in (0.002, 0.005, 0.01, 0.02):
    w_n = omega_d / np.sqrt(1.0 - true**2)
    system = signal.lti([1.0 / m, 0.0, 0.0], [1.0, 2.0 * true * w_n, w_n**2])
    _, response, _ = signal.lsim(system, force, t)
    a3, exact = [], []
    for ratio in ratios:
        result = vibration.impact_mobility(force, response, fs,
                                           exponential_decay_rate_per_s=ratio * omega_d,
                                           frequency_range_hz=(5.0, 800.0))
        c = result.fit_mode((40.0, 60.0)).correction
        a3.append(float(c.damping_ratio[0]))
        exact.append(float(c.exact_damping_ratio[0]))
    plt.plot(100 * ratios, 100 * (np.array(a3) - true) / true, "o-",
             label=f"Formula (A.3), true damping {100 * true:g} %")
    plt.plot(100 * ratios, 100 * (np.array(exact) - true) / true, "k:")
plt.xlabel(r"Damping the window adds, $a/\omega_r$ [%]")
plt.ylabel("Error in the true damping [%]")
plt.legend(); plt.show()
```

</details>

Two practical points close the annex. The window is often applied to the
response alone: the pulse is so short that windowing it amounts to multiplying
the force by the window's value at its peak, which `exponential_on_force=False`
does, and which is close to 1 when the pulse sits near the start of the
record. And if the calibration of 7.2 uses the same window and pre-trigger as
the measurement, that factor is already in the calibration.

## 5. Checks on the record

Every check returns a verdict object with `passes` and a `.plot()`; reading one
as a truth value raises, so a verdict is never mistaken for a number.

**Double hits (6.4).** Two impacts in one record are two pulses whose
transforms interfere: a second pulse $r$ times the first and $\tau$ later
multiplies the spectrum by $1 + r \mathrm{e}^{-j\omega\tau}$, which ripples between
$(1 + r)^2$ and $(1 - r)^2$ every $1/\tau$ hertz. The notches are deep only
when the second impact is about as strong as the first, which is why the part
says a small second impact shows as a slight ripple and moderate dips may be
tolerated. At a notch there is almost no force, and the mobility there is
noise.

```python
double = force + 0.6 * np.roll(force, round(0.06 * fs))   # 60 % again, 60 ms later
hit = vibration.check_double_hit(double, fs)
print(hit.passes, hit.impacts, round(hit.delay_s, 3))     # False 2 0.06
print(round(hit.notch_spacing_hz, 1), round(hit.ripple_db, 1))  # 16.7 12.0
print(vibration.check_double_hit(force, fs).passes)       # True
```

An impact is a run of samples above a fraction of the largest force, 0.1 by
default (`threshold_ratio`); that number is this library's, since the part
gives none. At 0.1 an undetected second pulse of the same shape as the first
ripples the spectrum by at most 1.74 dB. A softer, longer rebound, the usual
hammer bounce, carries more of the low-frequency spectrum than its peak says,
and can stay under the threshold while it ripples the spectrum by several
decibels, which is why 6.4 finds multiple impacts most easily in the frequency
domain, where the force-spectrum check below reads them. Run the check on the
unwindowed force, and on the unfiltered one when the analyser offers it: the
anti-aliasing filter can hide a small second impact in the ringing of the
first, which is why 6.4 recommends a storage oscilloscope on the unfiltered
signal, and a steep filter can ring above the threshold itself, so that one
impact reads as several.

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/impact_double_hit_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/impact_double_hit.svg" alt="Two panels. Left, a force record over 100 milliseconds with two Gaussian impacts marked, about 97 newtons at 5 milliseconds and about 58 newtons at 65 milliseconds. Right, the force energy spectral density up to 400 hertz on a logarithmic scale: a single impact falls smoothly from about 0.03, and the double hit oscillates around it, rising to about two and a half times and dipping to about a sixth of it every 16.7 hertz." width="92%"></picture>

*A double hit in the force record. The two impacts are 60 ms apart, so the
spectrum dips every 16.7 Hz; with the second at 60 % of the first, it swings
between 2.56 and 0.16 times the single pulse's, 12 dB peak to notch.*

<details>
<summary>Show the code for this figure</summary>

`DoubleHitCheck.plot()` draws the record with its impacts and the spectrum
with its notches:

```python
hit.plot()
```

The figure itself, by hand:

```python
import matplotlib.pyplot as plt

fig, (left, right) = plt.subplots(1, 2, figsize=(12.4, 5.2))
shown = t <= 0.1
left.plot(1e3 * t[shown], double[shown], label="force record")
left.plot(1e3 * hit.impact_times_s, hit.impact_peaks, "o",
          label="two impacts, 60 ms apart")
left.set_xlabel("Time [ms]"); left.set_ylabel("Force [N]"); left.legend()

_, one = vibration.energy_spectral_density(force, fs)
_, both = vibration.energy_spectral_density(double, fs)
band = (f > 0.0) & (f <= 400.0)
right.semilogy(f[band], one[band], "--", color="0.6", label="one impact")
right.semilogy(f[band], both[band], label="both: dips every 16.7 Hz")
right.set_xlabel("Frequency [Hz]"); right.set_ylabel("$G_{FF}$ [N²·s/Hz]")
right.legend(); plt.show()
```

</details>

**The force spectrum (6.2, 6.3).** The pulse should put its energy into the
frequency range of interest and not far beyond it. The check reads how far the
energy spectral density falls across the range, from its highest to its lowest
value there, against a limit the caller states, because the part prints none;
beside it, the share of the force energy above the range, which 6.3 asks to
keep small:

```python
spectrum = vibration.check_force_spectrum(force, fs, frequency_range_hz=(0.0, 400.0),
                                          max_drop_db=10.0)
print(spectrum.passes, round(spectrum.drop_db, 2))       # True 6.86 (dB)
print(round(100 * spectrum.energy_fraction_above, 1))    # 7.5 % above 400 Hz

wide = vibration.check_force_spectrum(force, fs, frequency_range_hz=(0.0, 1000.0),
                                      max_drop_db=10.0)
print(wide.passes, round(wide.drop_db, 1))               # False 42.9: a harder tip
```

A double hit fails the same check through its notches (18.4 dB across 0 to
400 Hz for the record above), which is the part's own advice: multiple impacts
are most easily detected in the frequency domain (6.4).

**Overload (8.4).** Impact signals carry energy out of band, and the dynamic
range is used to its limit, so saturation is a real risk. The check compares
each channel with the manufacturer's maximum for linear operation; a pass is
necessary, not sufficient, because a filtered digital record gives poor
definition of the waveform the converter saw:

```python
over = vibration.check_overload(accel, fs, full_scale=50.0)
print(over.passes, round(over.peak, 1), round(over.headroom_db, 1))   # True 47.6 0.4
```

**Channel match (8.1).** One broad-band signal into both channels, the
frequency response between them estimated, for instance with
`phonometry.electroacoustics.frequency_response.transfer_function`, and
`verify_channel_match(frequencies, response, frequency_range_hz=...)` asks
for unity within ±5 % and zero phase within ±5° across the range. It is the
one check here that judges the instrument rather than the record, hence
`verify_` rather than `check_`.

**Calibration (7.2).** The rigid-block check of ISO 7626-2 7.5.2,
`rigid_mass_calibration_check`, described on the [mobility
page](https://docs.jmrp.io/phonometry/vibration/structural/mechanical-mobility/). The block's mass
shall be chosen so its mobility is representative of the mobilities to be
measured, and the hammer and the analyser shall be set up exactly as in the
measurement.

## 6. The fiche

`ImpactMobilityResult.report(path)` renders the one-page ISO 7626 fiche with
the basis line naming ISO 7626-5:2019 and the averaging of 8.6, and the table
extended by the number of impacts, the exponential window and the coherence at
the peak. When a window was used, the fiche says the peaks carry its damping:
the corrected mobility is a modal result, and the fiche reports the
measurement.

```python
from phonometry import ReportMetadata

windowed.report(
    "impact_mobility.pdf",
    metadata=ReportMetadata(
        specimen="Machine support bracket (driving point)",
        measurement_standard="ISO 7626-5",
    ),
)   # one-page fiche (needs phonometry[report,plot])
```

[![ISO 7626-5 impact-mobility example report: a metadata header, a table of the characteristic points (the FRF type, the frequency range, the peak frequency and mobility, the phase there, the impacts averaged, the exponential window and the coherence at the peak) beside the averaged mobility spectrum, and the boxed peak mobility with a note that the peaks carry the window's damping](https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/reports/iso7626_5_impact_mobility_example.webp)](https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/reports/iso7626_5_impact_mobility_example.pdf)

## See also

- [Mechanical mobility and the FRF family (ISO 7626-1)](https://docs.jmrp.io/phonometry/vibration/structural/mechanical-mobility/):
  the FRF vocabulary this page estimates in, the attached-exciter measurement
  of ISO 7626-2 and the rigid-mass calibration.
- [Frequency response and coherence](https://docs.jmrp.io/phonometry/devices/electroacoustics/electroacoustics/):
  the `transfer_function` estimator for the channel-match test of 8.1.
- [Transfer stiffness of resilient elements (ISO 10846)](https://docs.jmrp.io/phonometry/vibration/structural/transfer-stiffness/):
  a force-per-motion FRF measured on a laboratory rig.
- API reference: [`vibration.structural.impact_mobility`](https://docs.jmrp.io/phonometry/reference/api/vibration/impact-mobility/).

## References

- International Organization for Standardization. (2019). *Mechanical
  vibration and shock — Experimental determination of mechanical mobility —
  Part 5: Measurements using impact excitation with an exciter which is not
  attached to the structure* (ISO 7626-5:2019).
  The whole page: windows, averaging, coherence, the checks on the record and
  the correction of Annex A.
- International Organization for Standardization. (2015). *Mechanical
  vibration and shock — Experimental determination of mechanical mobility —
  Part 2: Measurements using single-point translation excitation with an
  attached vibration exciter* (ISO 7626-2:2015).
  [iso.org catalogue](https://www.iso.org/standard/62483.html).
  The rigid-mass calibration of 7.5.2 and the random-error formula of Annex A.
- International Organization for Standardization. (2011). *Mechanical
  vibration and shock — Experimental determination of mechanical mobility —
  Part 1: Basic terms and definitions, and transducer specifications*
  (ISO 7626-1:2011).
  [iso.org catalogue](https://www.iso.org/standard/50426.html).
  The FRF family the estimate is expressed in.
- International Organization for Standardization. (2018). *Mechanical
  vibration, shock and condition monitoring — Vocabulary* (ISO 2041:2018).
  The damping ratio, the damping coefficient over the critical one (3.2.96),
  against which Formula (A.3) is first order; ISO 7626-5 takes its terms from
  it (clause 3).

## Standards

ISO 7626-5:2019, *Mechanical vibration and shock — Experimental determination
of mechanical mobility — Part 5: Measurements using impact excitation with an
exciter which is not attached to the structure*: the energy spectral density of
3.3 and 3.4, the double hits of 6.4, the calibration of 7.2, the channel match
of 8.1, the sampling and decay rules of 8.3, saturation (8.4), the force window
of 8.5.1, the exponential window of 8.5.2, the averaging of 8.6, the coherence
of 9.1 and the correction of Annex A, Formulas (A.1) to (A.3). The part prints
no worked example, so conformance is anchored on closed forms: a resonator
struck by a Gaussian pulse, whose response is known exactly, gives its mobility
back from the averaged estimate, and its damping and mobility back after
exponential windowing and the Annex A correction. ISO 7626-2:2015 supplies the
rigid-mass calibration that 7.2 refers to.