# Errata in IEC 61672

> One page of the phonometry documentation, as Markdown. Index: https://docs.jmrp.io/phonometry/llms.txt

Canonical: https://docs.jmrp.io/phonometry/reference/errata/iec-61672/
Language: en
Alternate: https://docs.jmrp.io/phonometry/es/reference/errata/iec-61672/index.md
Version: phonometry 4.0.0
Install: pip install "phonometry[full]"
Author: José Manuel Requena Plens (https://orcid.org/0000-0003-1250-6212)
Published: 2026-10-28
Updated: 2026-11-16T21:36:14Z
License: MIT (https://github.com/jmrplens/phonometry/blob/main/LICENSE)
Summary: Errata in IEC 61672: 1 defect in the printed text, with the evidence and what phonometry does about it.

---

This is the entry of the [errata registry](https://docs.jmrp.io/phonometry/reference/errata/) on IEC 61672-1:2013. The registry's introduction says how each claim is checked against the printed page and what the status of an entry means.

## IEC 61672-1:2013, Table 3 (the class 1 lower limit at 6 300 Hz printed with a decimal point)

- **Location:** Table 3, "Frequency weightings and acceptance limits", row
  6 300 Hz, column "Performance class 1", printed p. 22.
- **The print:** "+1,5; -2.0".
- **The problem:** every other number of the table writes its decimals
  with a comma, as the IEC does: the class 2 cell of the
  same row reads "±4,5" and the class 1 cell of the next row "+1,5; -2,5".
  The lower limit at 6 300 Hz is the one cell written with a decimal point.
  The value is not in doubt, -2.0 dB, between the -1.5 dB above it and the
  -2.5 dB below it.
- **Evidence:** Verified on PDF page 24 (printed p. 22) of BS EN
  61672-1:2013, the UK implementation of IEC 61672-1:2013 (second edition);
  the cell is legible and reads "-2.0" beside "+1,5;".
- **Library behaviour:** none needed. The library holds the number,
  -2.0 dB, in [`filters.weighting_class_limits`](https://github.com/jmrplens/phonometry/blob/main/src/phonometry/filters/weighting_compliance.py),
  which [`verify_sound_level_meter_periodic`](https://github.com/jmrplens/phonometry/blob/main/src/phonometry/metrology/sound_level_meter.py)
  reads for the frequency weightings.
- **Status:** unreported.