Summary: Errata in IEC 60118: 9 defects in the printed text, with the evidence and what phonometry does about each.

---

These are the 9 entries of the [errata registry](https://docs.jmrp.io/phonometry/reference/errata/) on IEC 60118-4:2014 and IEC 60118-4:2014/A1:2017, in the order the registry lists them. The registry's introduction says how each claim is checked against the printed page and what the status of an entry means.

## IEC 60118-4:2014, 6.4 NOTE 2, and IEC 62489-1:2010, 5.4.8.2 NOTE 3 (the responses of the band-limiting filters exchanged)

- **Location:** IEC 60118-4:2014, 6.4 "Pink noise signal", NOTE 2 (printed
  folio 13); IEC 62489-1:2010+A1:2014, 5.4.8.2 b), NOTE 3 (printed folio 11),
  which repeats it for the compliance voltage signal.
- **The print:** "the theoretical responses of the specified 3rd order
  Butterworth filters are **−0,8 dB at 100 Hz** and **−0,7 dB at 5 kHz**".
- **The problem:** both clauses specify third-order Butterworth high-pass and
  low-pass filters "giving −3 dB responses at 75 Hz and 6,5 kHz". Their
  combined response is $10\lg\{[1 + (75/f)^6]^{-1}[1 + (f/6\,500)^6]^{-1}\}$
  dB. At 100 Hz the high-pass term is $(75/100)^6 = 0.178$ and the response
  $-0.711$ dB; at 5 kHz the low-pass term is $(5\,000/6\,500)^6 = 0.207$
  and the response $-0.818$ dB. To the tenth the note prints they read
  **−0.7 dB at 100 Hz** and **−0.8 dB at 5 kHz**: the two values are
  exchanged. The ±1 dB flatness the note explains holds either way.
- **Evidence:** the Butterworth magnitude evaluated at the two frequencies.
  Verified on PDF page 15 (printed p. 13) of BS EN 60118-4:2015, the English
  text of EN 60118-4:2015, which is IEC 60118-4:2014 unchanged, and on PDF
  page 13 (printed p. 11) of BS EN 62489-1:2010+A1:2015, the English text of
  EN 62489-1:2010+A1:2015, which is IEC 62489-1:2010 with its Amendment 1:2014
  unchanged.
- **Library behaviour:** `electroacoustics.band_limit_response` evaluates the
  two filters, and the conformance rows on 6.4 NOTE 2 pin $-0.711$ dB at
  100 Hz and $-0.818$ dB at 5 kHz with the printed values named as the
  erratum; `electroacoustics.loop_test_noise` filters with the same sections.
- **Status:** unreported.

## IEC 60118-4:2014, 6.4 (the filter order printed as "one-third-order")

- **Location:** 6.4 "Pink noise signal", second paragraph (printed folio 13).
- **The print:** "Bandwidth limitation shall be carried out by means of at
  least **one-third-order** Butterworth high pass and low pass filters giving
  −3 dB responses at 75 Hz and 6,5 kHz".
- **The problem:** a filter has no one-third order: the wording crosses the
  "third-octave-band spectrum" of the sentence before it with "third-order".
  The same document's B.2.3 reads "The band-limiting filters should be at
  least of the **third-order**", NOTE 2 to 6.4 speaks of "the specified **3rd
  order** Butterworth filters", and IEC 62489-1 5.4.8.2 b), which specifies
  the same signal, prints "at least **third-order** Butterworth". The clause
  reads "at least third-order".
- **Evidence:** the three passages compared. Verified on PDF page 15 (printed
  p. 13) and PDF page 32 (printed p. 30) of BS EN 60118-4:2015, the English
  text of EN 60118-4:2015, which is IEC 60118-4:2014 unchanged, and on PDF
  page 13 (printed p. 11) of BS EN 62489-1:2010+A1:2015.
- **Library behaviour:** `electroacoustics.loop_test_noise` and
  `electroacoustics.band_limit_response` use third-order Butterworth sections;
  no change was required.
- **Status:** unreported (typographic, no numerical consequence).

## IEC 60118-4:2014, 6.6 and 8.2.1 (the meter cross-referenced to 5.1)

- **Location:** 6.6 "Combi signal", third paragraph (printed folio 14), and
  8.2.1 "Characteristic to be specified" (printed folio 16).
- **The print:** 6.6: "to allow either meter specified in **5.1** to reach
  the correct measurement level"; 8.2.1: "measured with a meter as specified
  in **5.1** and a pick-up coil whose magnetic axis is vertical (unless
  otherwise specified – see **6.1**)".
- **The problem:** 5.1 is the "General" subclause of Clause 5, "Using
  components of a sound system in an induction-loop system" (printed folio
  11), and specifies no meter. The two meters are specified in 6.1, the
  true-RMS meter in 6.1.3 and the peak programme meter in 6.1.4, and the
  vertical pick-up coil "unless otherwise specified" is the rule of 8.1
  (printed folio 15), which 7.1 and 10.2.1 cite for the same words. The
  references read **6.1** and **8.1**.
- **Evidence:** the two references compared with the headings of Clauses 5,
  6 and 8. Verified on PDF pages 13, 16, 17 and 18 (printed pp. 11, 14, 15
  and 16) of BS EN 60118-4:2015, the English text of EN 60118-4:2015, which
  is IEC 60118-4:2014 unchanged.
- **Library behaviour:** `electroacoustics.field_strength_meter` is the
  true-RMS meter of 6.1.3, which 6.1.1 makes definitive in case of doubt; no
  change was required.
- **Status:** unreported (cross-reference defect, no numerical consequence).

## IEC 60118-4:2014, Figure 2 a) (the outer radius keyed as 200 mm)

- **Location:** Figure 2 "Measurement points for disabled refuge and similar
  call-points", key of a) "Magnetic field source of small dimensions"
  (printed folio 21).
- **The print:** "$l_2$ inner radius 300", "$l_3$ **outer radius 200**", with
  $l_3$ dimensioned in the drawing from the inner ring of points to the outer
  one.
- **The problem:** an outer radius of 200 mm would lie inside the inner
  radius of 300 mm. As drawn, $l_3$ is the distance between the two rings,
  so the outer radius is $l_2 + l_3 = 500$ mm, which is also how far the
  second row of b) sits from its reference line ($l_2 + l_3 = 300 + 200$ mm).
  Amendment 1:2017 redraws the key as "$l_2 + l_3$ Radio exterior: 500 mm".
- **Evidence:** the key compared with the dimension lines of the drawing and
  with b). Verified on PDF page 23 (printed p. 21) of BS EN 60118-4:2015, the
  English text of EN 60118-4:2015, which is IEC 60118-4:2014 unchanged, and on
  PDF page 7 (printed p. 7) of UNE-EN IEC 60118-4:2016/A1:2018, the Spanish
  edition of IEC 60118-4:2014/A1:2017.
- **Library behaviour:** `electroacoustics.small_volume_measurement_points`
  places the outer ring at 500 mm, as the amended key does, and the
  conformance row on Figures 2 and 3 checks both radii.
- **Status:** unreported (corrected by the issuing body in 2017).

## IEC 60118-4:2014, Annex A.4 against 9.5 (+12 dB at 1.45 m for a counter)

- **Location:** Annex A (informative), A.4 "Specific locations such as help
  and information points, ticket and bank counters", the two paragraphs after
  Figure A.2 (printed folio 29), against 9.5 "Requirements for counter
  systems" (printed folio 24).
- **The print:** A.4: "At this height, the field strength level ref.
  400 mA/m at some of the points may be up to, but not greater than,
  **+12 dB**", of the measurements at 1,45 m, and "For the points in the plan
  view, a **less stringent** requirement is appropriate for a counter
  system". 9.5: "The magnetic field strength level at these points shall be
  ±6 dB ref. 400 mA/m", at "all measurement points specified in Figure 3,
  across both vertical and horizontal ranges", and "The field strength shall
  not be above +8 dB ref. 400 mA/m in the area where people are expected to
  stand".
- **The problem:** Figure 3 puts the counter's points at 1.2 m, 1.45 m and
  1.7 m, and 9.5 holds all of them to the ±6 dB of a refuge (9.3), with the
  +8 dB ceiling in the standing area, excepting no height. Nothing in the
  normative clause is less stringent for a counter than for a refuge, and the
  informative annex allows 4 dB more at 1.45 m than 9.5 allows anywhere.
  Amendment 1:2017 rewrites clause 9 with the same limits in its 9.5 and does
  not amend Annex A.
- **Evidence:** the two passages compared. Verified on PDF pages 26 and 31
  (printed pp. 24 and 29) of BS EN 60118-4:2015, the English text of EN
  60118-4:2015, which is IEC 60118-4:2014 unchanged, and on PDF page 10
  (printed p. 10) of UNE-EN IEC 60118-4:2016/A1:2018, the Spanish edition of
  IEC 60118-4:2014/A1:2017.
- **Library behaviour:** `electroacoustics.verify_small_volume_system`
  applies 9.5 as amended, ±6 dB at every measurement point and +8 dB in the
  standing area, at every height; the conformance row on 9.5 pins both edges.
- **Status:** unreported.

## IEC 60118-4:2014, Figure E.2 a) (the vertical-field line drawn along the length)

- **Location:** Annex E, Figure E.2 "Strengths of the components of the
  magnetic field due to current in a horizontal rectangular loop at points in
  a plane above or below the loop plane", panel a) (printed folio 38), against
  the text of E.1 that introduces it (printed folio 36).
- **The print:** panel a) draws a loop "15 units by 10 units" with two dashed
  lines through its centre. The one along the 15-unit length is labelled
  "**Vertical** field strength measured along a line parallel to this and 1,2
  units above or below the loop plane", the one across the 10-unit width
  "**Horizontal** field strength measured along a line parallel to this". E.1
  says that "Figure E.2 shows the distribution of the vertical component
  **across** a loop" and that "the horizontal axis gives position as a
  percentage of the loop **width**".
- **The problem:** both curves of panel b) are the traverse across the
  10-unit width, as E.1 says, and neither is the traverse along the length.
  The field of the loop, 1.2 units above it and in decibels re the field at
  its centre in its plane, reads for the vertical component at 5 %, 7.5 %,
  10 % and 12.5 % of the dimension $-0.3$, $+1.1$, $+1.7$ and
  $+1.9$ dB across the width, and $+1.8$, $+2.4$, $+2.4$ and
  $+2.1$ dB along the length; the printed curve reads about $-0.8$,
  $+0.8$, $+1.5$ and $+1.7$ dB, with its peak near 12.5 % where the
  field across the width peaks, not near 8.5 % where the field along the length
  does. The horizontal component tells the same: at $-10$ % it is
  $-0.1$ dB across the width and $-4.0$ dB along the length, and the
  curve reads $-0.2$ dB. The vertical-field line of panel a) belongs across
  the width, on the line labelled for the horizontal field.
- **Evidence:** the Biot-Savart field of the four sides of the 15 by 10 loop
  evaluated along both lines, against the curves read off the page from
  their own gridlines. Verified on PDF pages 38 and 40 (printed pp. 36 and
  38) of BS EN 60118-4:2015, the English text of EN 60118-4:2015, which is
  IEC 60118-4:2014 unchanged.
- **Library behaviour:** `electroacoustics.rectangular_loop_field` computes
  the field at any point, and the conformance rows on Figure E.2 b) compare
  the traverse across the width with points read off both curves.
- **Status:** unreported.

## IEC 60118-4:2014, E.6 (the flux density of 1 A/m printed as 1.256 µT)

- **Location:** Annex E, E.6 "Magnetic units", the bullet on magnetic
  induction (printed folio 44).
- **The print:** "$B = \mu_0\mu_r H$, where $\mu_0$ is the permeability of
  free space ($4\pi \times 10^{-7}$ H/m) [...] the magnetic induction due to a
  field strength of 1 A/m is **1,256 µT**".
- **The problem:** $4\pi \times 10^{-7}\ \mathrm{H/m} \times 1\ \mathrm{A/m} =
  1.256\,64\ \mu\mathrm{T}$, which reads **1.257 µT** to the four figures
  printed; the print drops the last digit instead of rounding it. The field
  strength of 79.58 A/m that the same bullet gives for one gauss in air,
  $10^{-4}\ \mathrm{T}/\mu_0 = 79.577$ A/m, is rounded correctly.
- **Evidence:** $4\pi \times 10^{-7}$ evaluated. Verified on PDF page 46
  (printed p. 44) of BS EN 60118-4:2015, the English text of EN
  60118-4:2015, which is IEC 60118-4:2014 unchanged.
- **Library behaviour:** `electroacoustics.magnetic_flux_density` multiplies
  by $\mu_0 = 4\pi \times 10^{-7}$ H/m; the conformance row on E.6 compares it
  with the printed 1.256 µT within 0.001 µT and names the print as the
  erratum.
- **Status:** unreported (typographic, no numerical consequence).

## IEC 60118-4:2014/A1:2017, 8.4 (cross-references to 10.2.7 left behind by the renumbering)

- **Location:** 8.4.1, 8.4.2 and 8.4.3 of IEC 60118-4:2014 (printed folio
  19), which Amendment 1:2017 does not amend, against clause 10 as the
  amendment replaces it (printed folios 11 to 13 of the Spanish edition).
- **The print:** 8.4.1: "The volume within which the requirements
  recommended or specified in Clause 7, 8.2.7, 8.3.7 and **10.2.7** are met";
  8.4.2: "See Clause 7, 8.2, 8.3 and **10.2**"; 8.4.3: "the requirements in
  8.3.7 and **10.2.7** apply". In the 2014 text 10.2 is "Magnetic noise level
  due to the system" and 10.2.7 its requirement.
- **The problem:** Amendment 1 replaces the text of clause 10. Its 10.2 is
  the commissioning requirement ("Requisitos") with no 10.2.7, and the noise
  with the system switched on moves to 10.4, whose 10.4.7 carries the rule of
  the old 10.2.7 word for word. The amendment leaves 8.4 untouched, so after
  it the useful magnetic field volume is defined by a subclause that no longer
  exists. The references read **10.4.7** and **10.4**.
- **Evidence:** the headings of clause 10 in both texts compared with the
  references in 8.4. Verified on PDF pages 21 and 26 (printed pp. 19 and 24)
  of BS EN 60118-4:2015, the English text of EN 60118-4:2015, which is IEC
  60118-4:2014 unchanged, and on PDF pages 11 and 13 (printed pp. 11 and 13)
  of UNE-EN IEC 60118-4:2016/A1:2018, the Spanish edition of IEC
  60118-4:2014/A1:2017.
- **Library behaviour:** `electroacoustics.verify_induction_loop_system`
  judges the noise with the system switched on by 10.4.7 and names that
  subclause in its verdict; no change was required.
- **Status:** unreported (cross-reference defect, no numerical consequence).

## IEC 60118-4:2014, 10.2.7, and IEC 60118-4:2014/A1:2017, 10.4.7 (a reference signal-to-noise ratio of exactly 47 dB)

- **Location:** 10.2.7 "Requirements" of IEC 60118-4:2014 (printed folio
  25), and 10.4.7 "Requisitos" of Amendment 1:2017, which carries the same
  rule after the renumbering (printed folio 13 of the Spanish edition).
- **The print:** 10.2.7: "If the reference signal-to-noise ratio as measured
  in 7.2 is **greater than 47 dB**, the magnetic field strength level at any
  point with the system switched on shall not exceed −47 dB. If the reference
  signal-to-noise ratio is **less than 47 dB** then the magnetic field
  strength level at any point with the system switched on shall not exceed
  that with the system switched off by more than 1 dB." 10.4.7 of the
  amendment: "es mayor de 47 dB [...] es menor de 47 dB".
- **The problem:** a site whose reference signal-to-noise ratio is exactly
  47 dB falls in neither sentence, so the clause sets no limit on the noise
  with the system switched on there, and the two rules give different limits
  at that point: with the noisiest point at −47 dB the first caps the system
  noise at −47 dB, the second at −46 dB. 7.2 calls only a ratio "greater than
  47 dB" ideal, which puts 47 dB with the lower branch.
- **Evidence:** the two sentences of each edition compared. Verified on PDF
  page 27 (printed p. 25) of BS EN 60118-4:2015, the English text of EN
  60118-4:2015, which is IEC 60118-4:2014 unchanged, and on PDF page 13
  (printed p. 13) of UNE-EN IEC 60118-4:2016/A1:2018, the Spanish edition of
  IEC 60118-4:2014/A1:2017.
- **Library behaviour:** `electroacoustics.verify_induction_loop_system`
  applies the −47 dB ceiling only above 47 dB and the 1 dB rule at 47 dB and
  below, and says so in its docstring.
- **Status:** unreported.