Calibration utilities for mapping digital signals to physical SPL levels.

> Auto-generated from the source docstrings by `scripts/generate_api_docs.py` (`make api-docs`). Do not edit by hand.

## CalibrationWarning

The calibration reference recording looks unreliable.

## sensitivity

```python
sensitivity(
    ref_signal: Signal,
    target_spl: float = ...,
    reference_pressure_pa: float = ...,
    fs: int | None = ...,
    *,
    validate: bool = ...,
    max_fluctuation_db: float | None = ...,
    frequency: float = ...,
    calibrator_class: str = ...,
    narrowband: bool = ...,
) -> float

sensitivity(
    ref_signal: SignalInput,
    target_spl: float = ...,
    reference_pressure_pa: float = ...,
    *,
    fs: int,
    validate: bool = ...,
    max_fluctuation_db: float | None = ...,
    frequency: float = ...,
    calibrator_class: str = ...,
    narrowband: bool = ...,
) -> float

sensitivity(
    ref_signal: SignalInput,
    target_spl: float,
    reference_pressure_pa: float,
    fs: int,
    *,
    validate: bool = ...,
    max_fluctuation_db: float | None = ...,
    frequency: float = ...,
    calibrator_class: str = ...,
    narrowband: bool = ...,
) -> float

sensitivity(
    ref_signal: SignalInput,
    target_spl: float = ...,
    reference_pressure_pa: float = ...,
    fs: int | None = ...,
    *,
    validate: bool = ...,
    max_fluctuation_db: float | None = ...,
    frequency: float = ...,
    calibrator_class: str = ...,
    narrowband: Literal[False] = ...,
) -> float
```

Calculate the calibration factor (multiplier) to convert digital units
to Pascals based on a reference recording (e.g., 1kHz @ 94dB).

When `fs` is provided (and `validate` is True), the recording's
stability is checked the way IEC 60942:2017 specifies for the calibrator
itself (5.3.3): levels are measured with time-weighting F and the
*short-term level fluctuation* (the absolute difference between each of
the maximum and minimum levels and the mean level) must not exceed the
Table 2 acceptance limit for the calibrator's class and nominal frequency,
read from [`FLUCTUATION_ACCEPTANCE_LIMITS_DB`](/phonometry/reference/api/metrology/sound-calibrator/#fluctuation_acceptance_limits_db).
For class 1 that is 0.07 dB at and above 160 Hz, relaxed to 0.10 dB above
63 Hz and below 160 Hz, and to 0.20 dB for the 31.5-63 Hz row where the F
time-weighting itself ripples; class LS is held to 0.03 dB and class 2 to
0.15 dB. Where Table 2 gives the class no limit (outside 31.5 Hz to 16 kHz
for class 1, outside 160 Hz to 1.25 kHz for LS and 2) the strictest limit
of the class's column applies. A larger fluctuation usually means a badly
coupled microphone or handling noise in the recording, which would
silently corrupt every calibrated level; a [`CalibrationWarning`](/phonometry/reference/api/metrology/calibration/#calibrationwarning) is
emitted.

:::note
IEC 60942 certifies calibrators over 60 s of operation sampled
at least 30 times; this check applies the same criterion to whatever
settled portion of the recording is available (>= 1 s after the 1 s
integrator attack).
:::

**Parameters**

| Name | Description |
| :--- | :--- |
| `ref_signal` | Recording of the calibration tone. Accepts a [`phonometry.io.Signal`](/phonometry/reference/api/io/io/#signal) for its rate; a calibration factor it carries is deliberately **not** applied, because this function is what produces such a factor and folding an existing one in would calibrate the calibration. |
| `target_spl` | The known SPL level of the calibrator (default 94 dB). |
| `reference_pressure_pa` | Reference pressure (default 20 microPascals). |
| `fs` | Sample rate of the recording in Hz. Required for the stability validation; without it the check is skipped. A [`Signal`](/phonometry/reference/api/io/io/#signal) supplies it, so a read take gets the validation for free, and an explicit value that disagrees raises. |
| `validate` | If True (default) and `fs` is given, warn when the recording's short-term level fluctuation exceeds the limit. |
| `max_fluctuation_db` | Explicit fluctuation limit in dB. Default (None) resolves the IEC 60942:2017 Table 2 limit for `calibrator_class` at `frequency`. |
| `frequency` | Nominal frequency of the calibration tone in Hz (default 1000.0), used to select the Table 2 row. |
| `calibrator_class` | The calibrator's class designation, `"LS"`, `"LS/M"`, `"1"` (default), `"1/M"` or `"2"` (IEC 60942:2017 Table 1), used to select the Table 2 column. |
| `narrowband` | If True (requires the rate, as `fs` or carried by a [`Signal`](/phonometry/reference/api/io/io/#signal)), estimate the tone level with a coherent single-frequency (Goertzel) detector locked to the tone near `frequency` instead of the full-band RMS. This rejects broadband hum/noise in the reference take, which otherwise inflates the RMS and shrinks the factor by $-10 \log_{10}(1 + 1/\mathrm{SNR})$ (about -0.44 dB at 20 dB SNR), silently biasing every subsequent level. The default (False) keeps the exact legacy broadband-RMS behaviour; enable it for noisy coupler recordings. |

**Returns:** Calibration factor (sensitivity multiplier).